Operating Specifications
|
| Analysis Modes |
Standard:TC, NPOC Option: TIC, TNb , TOCdiff, POC/VOC, |
| Ranges |
TC/Carbon : 0-10; 0-200; 0-2,000; 0-10,000; with Dilution 0-50,000 mg/L;
Wide Range: 0-200; 200-20,000 mg/L;
TNb*: 0-150 and 0-1,500 mg/L;
POC/VOC*: 0-25 mg/L; |
| Options/Accessories |
Leak Detection (for TC only), Sample Stream 2, Particulate Sampler, Air Box,
Air Scrubber, Communication Box, Mounting Rack, PID Detector* (for POC/VOC)
or ECD Detector* (for TNb) |
| Features |
Auto Calibration, Dilution, Check Function, Auto Rinse, Connectivity
(Remote Control), Predictive Maintenance |
| Precision |
+/-2% End of Detector Range† |
| Linearity |
R2 ≥0.997† |
| LOD |
<1 ppm† |
| Analysis Time |
3-5 min typically |
| Smallest Inner Diameter (ID) |
2 mm smallest tubing diameter |
| Capacity |
Up to 21 samples in duplicate with positive product controls |
| Salt Tolerance |
3-5% NaCl |
| Carrier Gas |
CO2 free air < 1 ppm CO2 and <0.1 ppm hydrocarbon: 100-200 kPa @ 30-50L/hr flow rate (Optional: Air box) |
| Sample Temperature |
2-60 °C (35.6-140 °F) |
| Ambient Temperature |
2-40 °C (35.6-104 °F) |
| Minimum Sample Flow Rate |
100 mL/min without Particulate Sampler or 11.7-50 LPM (3.1-13.2 US gal/min)
with Particulate Sampler |
| Drain |
Gravity, No back pressure (< 7 kPa, 0.07 Bar, 1 psig) |
Analyzer Specifications
|
| Communication protocols |
OPC UA, Modbus TCP/IP |
| Inputs |
Digital Input (4) |
| Display |
7 in. Touch panel |
| Power |
AC 110 – 230 V ±10 V 50/60 Hz, 400-600 VA |
| Alarm Relays |
Warnings, errors, measurement limit, instrument status |
| Outputs |
OPC UA, Ethernet, 4-20 mA (6) |
| Installation/Overvoltage Category |
II |
| Available Configurations |
Standard: Single-Stream; Option: Second Stream |
| Dimensions |
900 H x 644.3 W x 354.5 D mm (35.5 H x 25.4 W x 14 D in) |
| Weight |
55 kg, (121 lb) |
| Safety Certifications |
CE, UL, CSA |
| Regulatory |
DIN EN 1484 /ISO 8245, EPA 415.3, SM 5310B, ASTM D5173, HJ 501, DIN EN 12260 |
Environment
|
| Enclosure |
Standard: IP-54; Option: ATEX Zone 1 & 2 T4, IECEx Zone 1 & 2 T4, C1D2 |
| Maximum Relative Humidity |
Up to 85%, non-condensing |